The history of
microscopy is closely linked to the history of microbiology. Early microscopists,
including Hooke, Divini, Kircher, and van Leeuwenhoek, were among the first
individuals to describe microscopic life forms . Robert Hooke, in his landmark
1665 book Micrographica, included many illustrations of microscopic forms. In
observation XVIII, Hooke first used the term "cell" to describe the
microscopic architecture of thin pieces of cork. Hooke's Micrographica was a
best seller of its time, inspiring many scientists to discover the
microbiological world. In his 1678 letters to the Royal Society in London,
Antony van Leeuwenhoek provided detailed descriptions of protozoa. His
descriptions of "very small animalcules" included drawings of basic
organism shapes and descriptions of their movement. In an effort to observe the
microscopic world in ever greater detail, van Leeuwenhoek built simple
microscopes of increasing power. Some of these microscopes were capable of x200
magnification. Van Leeuwenhoek's 1683 letters to the Royal Society included the
first ever description of living bacteria from dental plaque sampled from
between his teeth. Microscopy continued to play an important role in the study
of biology and medicine, and almost two centuries later, the only scientific
equipment Charles Darwin took with him on the voyage of the Beagle was a simple
microscope. Today, light microscopy is used not only in biological sciences
such as microbiology, botany, forensics, pathology, and cell biology but also
in metallurgy, engraving, chemistry, mineralogy, gemology, computer chip
design, and microsurgical applications. This chapter will attempt to describe
the basic concepts of light microscopy as they are practiced in the
microbiology laboratory.
TECHNICAL
BACKGROUND AND DEFINITION OF TERMS
Aberration
Aberrations
are unwanted artifacts in the microscopic image that are caused by elements in
the optical path. Aberration can be caused by physical objects such as dust or
oils on the optical surfaces, by alterations in the light path caused by
improper alignment or aperture settings, and by imperfections in the lens
systems. Two main types of optical aberrations can occur when white light
passes through a convex lens: spherical aberration and chromatic aberration.
Spherical aberration is hallmarked by images that appear to be in focus in the
center of the field and out of focus at the periphery. Chromatic aberration
occurs because shorter light wavelengths are refracted to a greater extent than
longer wavelengths. This wavelength separation (also called dispersion)
produces color fringes within the image field. Chromatic aberration can be
reduced or eliminated in optical systems by combining two lenses with different
color dispersion characteristics.
Contrast
Contrast is a
measure of the differences in image luminance that provides gray scale or color
information. Contrast is expressed as the ratio of the difference in luminance
between two points divided by the average luminance in the field under optimum
conditions, the human eye can detect the presence of 2% contrast
Depth
of Field and Depth of Focus
Depth of field
is a subjective measure of the vertical distance between the nearest and
farthest objects in the specimen that appear to be in sharp focus. Depth of
field decreases as the numerical aperture (NA) of the lens increases Depth of
focus is the area around the image plane where the image will appear to be
sharply focused. The image plane is formed within the microscope tube at or
near the level of the ocular lenses. Microscopes with greater depth of focus
allow the user to employ ocular lenses with different working distances,
magnification factors, and visual compensation systems without losing image
sharpness. Like depth of field, depth of focus depends upon the NA of the
objective. However, depth of focus increases as the NA increases.
Immersion
Fluid (Immersion Oil) Immersion fluid is a term used to describe any liquid
that occupies the space between the object and microscope objective lens.
Immersion fluids are usually required for objectives that have working
distances of 3 mm or less. Many microscopy applications employ immersion fluids
that possess the same refractive index as the glass slide (refractive index
1.515). This procedure produces a homogeneous optical path which minimizes
light refraction and maximizes the effective NA of the objective lens.
Immersion fluids are also used between the condenser and the microscope slide
in transmitted light fluorescence microscopy and in dark-field microscopy to
minimize refraction, increase the NA of the objective, and improve optical
resolution.
Koehler Illumination
Kohler
illumination was first introduced in 1893 by August Kohler of the Carl Zeiss
Corporation as a method of providing the optimum specimen illumination In this
procedure, the collector lens projects an enlarged and focused image of the
lamp filament onto the plane of the aperture diaphragm. Because the light
source is not focused at the specimen, the specimen is bathed in a uniformly
bright, glare-free light that is not seriously affected by dust and imperfections
on the glass surfaces of the condenser. Koehler illumination is required to
produce maximum optical resolution and high-quality photomicrographs.
Mechanical
Tube Length
Mechanical
tube length describes the light path distance within the microscope body tube.
Tube length is measured from the objective opening in the nosepiece to the top
edge of the observation tube. Tube length is usually inscribed on the barrel of
the objective as the length in millimeters (e.g., 160, 170, 210, etc.) for
fixed lengths or the infinity symbol (co) for infinity-corrected tube lengths.
Many of the newer objectives are infinity corrected, while older objectives
will be corrected for 160-mm (Nikon, Olympus, and Zeiss) or 170-mm (Leica) tube
lengths Differences in mechanical tube length are among the major reasons why
objectives from one manufacturer cannot be used on a different
Numerical
Aperture
NA is a
measure of the r gilt-gathering capability of a lens or condenser. Higher-NA
objectives have better resolving power and brighter images than lower-NA
objectives.
Higher-NA
objectives also have shallower depths of field. The equation for determining NA
is given by NA n, sine, where n is the refractive index of the imaging medium
between the objective and the specimen and E is one-half the angular aperture
of the objective (HL2) (3, 5, g, 15). For visible light microscopy using
standard immersion oils, the theoretical maximum for NA is about 1.52. In
practice, however, a lens cannot accept an 1800 cone of light, so B has to be
slightly less. This results in a maximum practical NA of about 1.40.
Refractive
Index (Index of Refraction)
The index of
refraction is the ratio of the velocity of light in a vacuum to its velocity in
a transparent or translucent medium. As the refractive index of a material
increases, light beams entering or leaving a material are deflected (refracted)
to a greater extent. The refractive index of a medium depends upon the
wavelength of light passing through it. Light beams containing multiple
wavelengths (e.g., white light) are dispersed when they move into a different
medium because each wavelength in the beam is refracted to a slightly different
degree. Light dispersion causes chromatic aberration in microscope objectives
Refractive index is also an important variable in calculating NA (see
"Numerical Aperture" above). Moving from a high-dry microscope
objective that uses air as the imaging medium (refractive index of air 1.003)
to an oil immersion objective of the same power (refractive index of immersion
oil 1.515) increases the maximum theoretical NA of a given lens from 1.0 to
1.5, producing a 50% increase in light-gathering capability
Resolution
(Resolving Power)
The resolution
of an optical microscope is defined as the shortest distance between two points
that can be distinguished by the observer or camera system as separate entities.
The resolving power of a microscope is the most important feature of the optical
system because it defines our ability to distinguish fine details in a
specimen. The theoretical limit of resolution for a given lens is defined
mathematically as r K/(2NA), where r is the resolution, K is the imaging
wavelength, and NA is the numerical aperture of the lens. From this equation,
it is obvious that only the light wavelength and NA directly affect the
resolving power. Thus, a 40x oil objective with an NA of 1.30 can have the same
resolving power as a 100x oil 0b objective. In the same manner, the resolving
power of a 100x oil objective is higher when using UV wavelengths than it is
when using visible light.
Working
distance
Working
distance is the distance between the objective front lens and the top of the
cover glass when the specimen is in focus.The working distance of an objective
generally decreases as magnification increases The working distance of an
objective may not be inscribed on the barrel of older objectives, but newer
objectives often contain the working distance in millimeters.
Longer-working-distance objectives are important when examining the inside
surface of glass tubes (tube cultures) and cell culture flasks.
SIMPLE MICROSCOPES
Common objects
such as jewelers' loupes, handheld photographic slide viewers, and simple
magnifying or reading glasses are all examples of simple microscopes in routine
use today. A simple microscope is composed of a single biconvex magnifying lens
which is thicker in the center than at the periphery. In contrast with compound
microscopes, simple microscopes produce a magnified image that is in the same
orientation as the original object. Because of their low NA, simple microscopes
have limited resolution and magnifying power. Most commercial magnifiers are
able to produce an x2 to x30 magnification, and the better lenses will have a
resolution of about 10 um. Simple magnifiers are useful for dissection,
examination of bacterial colonies, and interpretation of agglutination
reactions.
COMPOUND
MICROSCOPES
The first
compound microscopes were constructed around 1590 by Dutch spectacle makers,
Zaccharias Janssen and Hans Janssen. The Janssen microscope consisted of an
object lens (objective) that was placed close to the specimen and the eye, or
ocular, lens that was placed close to the eye. The lenses were separated by a
body tube. In this microscope, the objective lens projected a magnified image
into the body tube and the eyepiece magnified the projected image, thereby
producing a two-stage magnification. Modern compound microscopes still use this
general design and have two separate lens systems mounted at opposite ends of a
body tube.
The
stereoscopic microscope combines two compound microscopes which produce
separate images for each eye. Stereoscopic microscopes may have one or two
separate objectives, and many have a zoom magnification function. These
microscopes are used for reflected or transmitted illumination, but the absence
of a substage condenser limits their NA and resolution. Stereomicroscopes are
useful in examining colonial morphology of bacteria, fungi, and cell cultures.
The modern
light microscope is composed of optical and mechanical components that,
together with the mounted specimen, make up the optical train. The optical
train of a typical bright-field microscope consists of an illuminator (light
source and collector lens), a substage condenser, specimen, objective,
eyepiece, and a detector. The detector can be a camera or the observer's eye.
Specimen
illumination is one of the most critical elements in optical microscopy.
Inadequate or improper sample illumination can reduce contrast in the specimen
and significantly decrease the resolving power of any microscope. There are
numerous commercially available illuminators for microscopes, but 50- or
100-watt tungsten halogen lamp systems are frequently used due to their low
cost and long life. Light generated by the light source is passed through a
collector and a field lens (Fig. 3) before being directed into the substage
condenser and onto the specimen. Image-forming light rays are captured by the
microscope objective and passed into the eyepieces or a camera port. Alignment
of the optical components of a microscope is critical to produce a good image.
The field diaphragm
The field diaphragm
is located in the light path between the light source and the substage
condenser. This iris- like mechanism controls the width of the light beam that
enters the substage condenser. The field diaphragm does not affect the optical
resolution, NA, or intensity of Illumination. However, the field diaphragm
should be centered in the optical path and opened far enough that it just
overfills the field of view. This adjustment is important for preventing glare
and loss of contrast in the observed image. When the field diaphragm is opened
too far, scattered light and reflections can degrade image quality.
Substage
Condenser
The substage
condenser is typically mounted beneath the microscope stage in a bracket that
can be raised or lowered independently of the stage. The substage condenser
gathers light from the field diaphragm and concentrates it into a cone of light
that illuminates the specimen with uniform intensity over the entire field of
view.
Adjustment of
the substage condenser is probably the most critical element for achieving
proper illumination, and it is the main source of image degradation and poor
quality photomicrography. The condenser light cone must be properly adjusted to
optimize the intensity and angle of light entering the objective. Because each
objective has different light-gathering capabilities (NA), the substage
condenser should be adjusted to provide a light cone that matches the NA of the
new objective. This is done by adjusting the aperture (or condenser) diaphragm
control. Substage condensers on newer microscopes have a scale embossed on the
condenser and an index mark on the aperture control that allows the user to
quickly switch from one NA range to another. Many manufacturers are now
synchronizing the NA gradations to correspond with the approximate NA of the
objectives.
In clinical
laboratory practice, the condenser aperture is often made smaller to improve
the contrast of wet mounts and some stained preparations. The condenser is
sometimes moved downward for the same purpose. These practices are effective
for a few applications, but they will result in decreased optical resolution.
Specimen illumination intensity should not be adjusted by opening and closing
the condenser diaphragm or by moving the condenser laterally in the light path.
Illumination intensity should be controlled through the use of neutral density
filters placed into the light path or by reducing the amp voltage.
Reducing the
voltage, however, will also alter the color of the incoming light, and voltage changes
are not recommended for photomicroscopy.
Objectives
The objective
lens is the most important single determinant of the quality of the image
produced by a particular microscope. When choosing a microscope, the purchaser
must select the magnification factor, NA, and the level of correction for each
objective. Lenses with higher NA values will have higher resolution and produce
a brighter field of view. The level of optical correction in the objective will
depend upon the ultimate use of the microscope.
Achromatic
(achromat) objectives are the least expensive objectives found on laboratory
microscopes. Achromat objectives are corrected for axial chromatic aberration
in two wavelengths (red and blue), and they are corrected for spherical aberration
in one color (green). The limited correction of achromatic objectives can cause
a number of optical artifacts when specimens are examined and photographed in
color (e.g., green images often have a reddish-magenta halo). Achromat
objectives produce the best results with light passed through a green filter
and when performing black and white photomicroscopy.
Flatness of
field is also a problem when using straight achromat objectives because the
center of the field is in focus while the edges are out of focus. In the past
few years, most manufacturers have begun providing flat-field corrections for
achromat objectives.
These
objectives are called plan-achromats.
The next
higher level of correction and cost is found in objectives called fluorites or
semiapochromats. Fluorite objectives are produced from advanced glass formulations
that allow for greatly improved correction of optical aberration. Like the
achromat objectives, the fluorite objectives are corrected chromatically for
red and blue light. unlike achromats, fluorites are corrected spherically for
two or three colors instead of a single color.
The superior
correction of fluorite objectives compared to achromats enables these
objectives to be made with a higher NA. Fluorite lenses therefore produce
brighter images than achromats. Fluorite objectives have better resolving power
than achromats and provide a higher degree of contrast, making them better
suited for color photomicrography in white light.
Apochromats
Apochromats
are the most highly corrected microscope lenses and the most costly.
Apochromats are corrected chromatically for three colors (red, green, and
blue), almost eliminating chromatic aberration, and are corrected spherically
for either two or three wavelengths. Apochromat objectives are the best choice
for color photomicrography in white light. Because of their high level of
correction, apochromat objectives usually have, for a given magnification,
higher NAs than do achromats or fluorites.
Fluorescence
Objectives
Fluorescence
objectives are designed with quartz and other special glasses that have high
transmission rates for IJV, visible, and infrared light. These objectives are
extremely low in autofluorescence, and they use specialized optical cements and
antireflection coatings that protect the lens and allow it to operate with a
wide variety of excitation wavelengths.
Correction for
optical aberration and NA values in UV fluor objectives usually approaches that
of apochromats, which contributes to image brightness and enhanced image
resolution.
The primary
drawback of high-performance fluorescence objectives is that many are not
corrected for field curvature and produce images that do not have uniform focus
throughout the entire field of view. This is not a large problem when
performing direct or indirect fluorescent antibody testing, but it can be
troublesome if you have to use the same objectives for bright-field or
phase-contrast microscopy. Microscope objectives that use air as the medium
between the coverslip and the objective lens are considered dry objectives. The
maximum working NA of a dry objective system is limited to 0.95. Higher NA
values can only be achieved using optics designed for immersion media.
Immersion media have the same refractive index and dispersion values as glass
(refractive index — 1.51). The use of immersion media produces a homogeneous
light path from the coverslip to the lens so that light is not refracted away
from the objective.
The use of
immersion fluids and immersion lenses significantly increases the NA and the
optical resolution of the system. In addition to "oil" lenses,
specially corrected objective lenses designed for glycerin and water immersion
are commercially available. The proper immersion fluid type is always stamped
on the side of the objective. The advantages of oil immersion objectives are
severely compromised if the wrong immersion fluid is utilized. Microscope
manufacturers produce immersion objectives with tight refractive index and
dispersion tolerances.
It is
therefore advisable to use only the immersion fluid recommended by the
objective
manufacturer. Mixing of immersion fluids from different manufacturers should be
avoided because mixing can produce unexpected crystallization artifacts or
phase separations that compromise image quality. Many high-power (NA, 20.8) dry
objectives are engineered to operate through 0.17-mm coverslips (designated as
number 1%). In practice, however, the total thickness of the specimen-coverslip
sandwich can be greater or less than 0.17 mm due to variations in coverslip
and/or mounting fluid thickness .
Under these
conditions, there will be noticeable spherical aberration in the microscopic
image. A 0.2-mm deviation in coverslip thickness will produce an 8% decrease in
image intensity when using a 0.79 NA objective and 57% with a 0.85 NA high-dry objective.
Therefore, some of the more
advanced dry
objectives are engineered with a coverslip correction collar that adjusts the
objective lens elements to compensate for coverslip thickness variations.
Objectives with a coverslip correction collar are labeled Corr, w/Corr, or CR.
However, this labeling is usually unnecessary because the objective has a
distinctive knurled ring and graduated scale on the side. The expected
coverslip thickness for an objective is etched on the barrel of the objective.
Eyepiece
or Ocular Objective
The eyepiece
or ocular objective contains the final lens system in the optical train. The
purpose of the ocular objective is to magnify and focus the projected image
onto the eye of the viewer. Ocular lenses generally have a magnification factor
of x 10 to x20, and the total magnification of the microscope is the product of
the objective magnification and the ocular magnification. Thus, a microscope
with a 40x objective and a lox ocular lens have a magnification value of x400.
Many eyepieces have a shelf at the level of the fixed eyepiece diaphragm that
allows for the insertion of ocular micrometers, pointers, or crosshairs. This
shelf is located at the focal plane of the image projected by the objective
lens so that the inserted element is in focus when the specimen image is in
focus.
DARK-FIELD
MICROSCOPY
Dark-field
microscopy is a specialized illumination technique that is used in the clinical
laboratory to detect thin organisms such as spirochetes and Leptospira.
High-resolution dark-field microscopy utilizes a specialized high-NA cardioid
dark-field condenser that blocks the central light path light and produces a
hollow cone of illumination that is directed away from the objective lens at an
oblique angle. Bacteria on the slide have a slightly different refractive index
than the surrounding medium, and light rays passing through the organism are
refracted into the objective lens. Light rays that do not pass through an
organism do not enter the lens. This type of illumination produces bright
organism profiles against a dark background. Dark-field microscopy requires
careful alignment of the condenser and the placement of immersion oil between
the slide and the sub stage condenser. Dark-field microscopy, when done
correctly, increases the resolution of the microscope to 0.1 um or less. The
resolution of bright-field microscopy is 0.2 u.
PHASE-CONTRAST
MICROSCOPY
In the early
1930s, the Dutch physicist F. Zernike introduced a new method for testing
optical systems which he called phase contrast. Originally designed as a
macroscopic testing system, the use of phase-contrast techniques in microscopic
systems allowed the user to examine unstained biological specimens that were
virtually transparent when observed under bright-field illumination. Zernike
was awarded the Nobel Prize in 1953 for his contributions to microscopy.
Phase contrast
is essentially an interference method wherein a ring annulus is placed directly
under the lower lens of the condenser to produce a hollow cylinder of light.
This light is essentially unchanged as it passes into the objective, and it
arrives at the rear focal plane of the objective in the shape of a ring. Light
that goes through the specimen is refracted and slowed slightly so that it is
out of phase by about one-quarter wavelength from the unchanged light. This
out-of-phase light is spread over the entire focal plane. Light passing through
the rear focal plane of the objective interacts with a separate ring-shaped
phase plate that alters the direct light path by another one-quarter wavelength
when the direct light and the refracted light arrive at the image plane; they
are out of phase by one-half wavelength. This out-of-phase light interacts
destructively so that specimen details appear as dark areas against a lighter
background.
Because the
phase-shifting calculations are based on a one-quarter wavelength of green
light, the phase image has the best resolution when a green filter is placed in
the light path. Green filters also allow the microscopist to use less-expensive
achromat lenses that are spherically corrected for green light. Phase
microscopy is an important tool for examining living and/or unstained material
in wet mounts
and cell
cultures.
However,
phase-contrast microscopy has lower resolution than bright-field microscopy of
stained specimens In addition, viewed objects are often surrounded by halos
that can obscure boundary details. Phase-contrast microscopy does not work well
with thick specimens because the phase shift may be greater than the expected
one-quarter wavelength. In phase-contrast microscopy, structures within living
cells appear as hills or craters, depending upon their optical thickness. This
pseudo-three-dimensional effect can greatly enhance image contrast in unstained
specimens.
FLUORESCENCE
MICROSCOPY
The
fluorescence microscope was developed in the early 1900s, and many of the
initial microscopic studies involved identification and localization of
compounds that autofluoresced when irradiated with UV light. In the lg30s, a
number of investigators began using fluorescent compounds to identify specific
tissue components and infectious agents that did not autofluoresce These stains
are not organism specific, but rather, they bind to and stain specific
structures within the organism. Examples of this type of staining include
acridine orange, auramine-rhodamine, calcofluor white, Evans blue, and Hoechst
33258. The use of fluorochrome-antibody conjugates (immunofluorescence) was
first described in the lg40s when Coons et al used fluorescein-labeled
antibodies to detect pneumococcal polysaccharide antigens in tissue sections of
infected mice.
Fluorescent
antibody staining expanded significantly with the development of fluorescein
isocyanate in 1950 (13) and the more stable fluorescein isothiocyanate (FITC)
derivative in 1958. Today, fluorescence microscopy is also used in conjunction
with nucleic acid hybridization to visualize the location of fluorescent in
situ hybridization and multicolor fluorescent in situ hybridization probes .
Fluorescence
microscopy is dependent upon the ability of fluorescent substances to absorb
light of a certain wavelength and emit light at a longer wavelength.
Fluorescence microscopes used for clinical microbiology most often utilize
fluorochromes that absorb near- UV light energy and reemit that light at a
lower, green or yellow wavelength. To work properly, the fluorescence
microscope must irradiate the specimen with I-IV excitation ight and separate
the much weaker emitted light from the brighter excitation light so that only the
emitted light reaches the eye. The resulting image consists of brightly shining
areas against a dark background. Older fluorescent microscopes are configured
for dark-field illumination. In these instruments, LIV excitation light enters
a dark-field condenser and the light is directed onto the specimen at an
oblique angle. Fluorescent compounds in the specimen absorb the excitation
light, and the emitted light is collected by the objective lens.
The emitted
light then passes through a barrier filter to remove any excitation light that
may enter the objective. These microscopes are difficult to use for routine
diagnostics because the condenser and the objective must be carefully oiled.
The dark-field condenser also reduces the effective NA of the objectives,
thereby producing dim images that lack resolution. Most modern fluorescence
microscopes use reflected light (epifluorescence). In these instruments, the
excitation light is directed downward through the objective and onto the
specimen. The emitted light and the reflected excitation light are collected by
the objective, and they pass through a dichoric mirror which removes the
excitation light and allows the longer- wavelength emitted light to form an
image.
With
epifluorescence, the objective acts as a condenser and the alignment and oiling
issues associated with a dark- field condenser are eliminated The visual field
is brighter with epifluorescence, the resolution is higher, and fluorescence
quenching only occurs in the field of view. Fluorescence microscopy requires
high levels of illumination because the quantum yield of most fluorochromes is
low. The most common lamps are mercury vapor (HBO) lamps, ranging in wattage
from 50 to 200 watts, or xenon vapor (XBO) lamps, which range from 75
to 200 watts.
It should be noted that lamp wattage is not necessarily a measure of usable
brightness in a fluorescence lamp. The HBO 100-watt lamp is 4 times brighter
than the 200- watt HBO lamp and 11 times brighter than the XBO 150-watt lamp.
When purchasing a fluorescence microscope, it is also important to determine
whether the emission spectrum of the lamp is compatible with the fluorochromes
used in the laboratory.
HBO and XBO lamps are under high pressure, and
care must be taken to prevent the lamps from exploding. One should never touch
these lamps with bare hands because oils on the fingers can etch or discolor
the glass. Fluorochromes must be excited by specific light wavelengths to
generate the maximum amount of emitted light. Therefore, specific exciter and
barrier filter combinations are used to maximize the quantum yield of the
fluorophore. Exciter filters are used to select the required light wavelengths
from the spectrum of light generated by the amp Excitation filters are provided
in narrow, medium, and wide band-pass configurations that pass a narrower and
wider range of light frequencies, respectively. Barrier filters block shorter
light wavelengths and allow longer wavelengths to pass through the filter.
Barrier filters are important because they remove the high-intensity excitation
light that could overwhelm the low-intensity emitted l' get. Barrier filters
also prevent I-IV light from entering the eye where it can cause cataracts and
retinal damage.
Wide band-pass
barrier filters general y produce brighter images, but care must be taken to
prevent the introduction of background right that could overwhelm the emitted
light. Epifluorescence microscopes also have a dichromatic mirror (beam
splitter) that reflects the incoming excitation light to the objective and
allows the emitted light to pass to the barrier filter and onto the objectives.
In most modern epifiuorescence microscopes, the barrier filter, excitation
mirror, and beam splitter are housed in removable optical blocks, and several
of these blocks can be installed in the microscope at one time. This
configuration allows the user to quickly change the excitation and barrier
filters to accommodate different fluorochromes. Care must be exercised when
selecting optical blocks. The excitation filter should match the excitation
wavelength of the fluorophore, and the emission barrier should allow the
emitted light to pass through. For example, direct fluorescent antibody testing
for viral antigens in cell smears typically employs FITC-labeled antibodies and
an Evans blue counterstain. Choosing an optical block with a 450- to
490-nm-pore-size excitation filter and a 515-nm-pore-size wide band-pass
barrier filter will produce a bright field of view, and the counterstained
cells will appear orange-red. By selecting a more restricted band-pass barrier
filter (520- to 560-nm pore size), the field of view will be darker and the red
emitted light from the Evans blue counterstain will not be visible. The images
produced by optical block will have more contrast because the background is
darker. Both filter combinations are appropriate for this task, but the final
choice will depend upon user preference.
One of the
major problems in the use and examination of fluorescent microscopic images is
the tendency of fluorophores to lose fluorescence when exposed to excitation
light for several minutes. This loss of fluorescence is caused by two
mechanisms: photobleaching and quenching. Photobleaching (fading) is a
permanent loss of fluorescence that is caused by chemical damage to the
fluorophore. Quenching is caused by the presence of free radicals, salts of
heave,' metals, or halogen compounds (2). Quench'ng can also be caused by
transfer of emission light energy to other fluorescent molecules in close
proximity to the fluorophore in a process called fluorescent resonance energy
transfer. To lessen the effect of Quenching, slides should be stored in the
dark at 2 to BOC.
In addition, the user should block the
excitation light path when not viewing or photographing the specimen. Most
epifiuorescence microscopes have a shutter in the light path for this purpose.
Quenching can be a significant problem when photographing fluorescent images
because the shutter may be open for a minute or more. Quenching can be reduced
somewhat by the addition of free radical scavengers such as p-phenylenediamine,
1,4-diazabicyclo(2,2,2)-octane (DABCO) (25), orn-propylgallate to the mounting fluid. p-phenylenediamine and
n- propy gallate can be used to reduce quenching in FITC and rhodamine. DABCO
is slightly less effective than p-phenylenediamine for FITC fluorescence, but
unlike p-phenylenediamine, DABCO does not darken when exposed to ight and it is
safer to use. Quench-resistant mounting media are also available from Vector
Laboratories (Burlingame, CA), Molecular Probes, Inc. (Carlsbad, CA), and
Bio-Rad Laboratories (Hercules, CA).
LINEAR
MEASUREMENTS (MICROMETRY)
The first
reported micrometric procedures were credited to Antony van Leeuwenhoek, who
used fine grains of sand as a gauge to determine the size of human
erythrocytes. Since then, a variety of methods have been used to determine the
dimensions of microscopic organisms. The crudest method for determining size in
the clinical laboratory involves comparing the object size to the measured or
calculated •Wew field size. Other micrometric techniques include the addition
of polystyrene beads of known size into the specimen.Comparative measurements
are then performed utilizing a photomicrograph or digital image. The accuracy
of this method is variable and depends on the homogeneity of the comparison
objects. Direct measurement of microorganisms can be done by placing them on
calibrated microscope slides or counting chambers. The accuracy of this method
depends on the separation distance between ruled lines but averages between 10
and 50 micrometers. The most common procedure used in the clinical laboratory,'
utilizes a graduated scale (reticle) located within one of the eyepieces.
Reticles must be calibrated against a stage micrometer for each objective. The
accuracy of this type of measurement is
approximately
2 to 10 micrometers (3 to 5%), depending on magnification and the resolution of
the stage micrometer.
PHOTOMICROSCOPY
Microscopists
began capturing microscopic images on film shortly after the photographic
process was invented (4). Micrographic images have long been used for
investigations of morphology, in scientific publications and lectures, and in
teaching. Modern film technologies have high resolution and clarity, but the
use of photomicrographs in day-to-day microscopy has been hampered by long
turnaround times associated with film development and printing. Reacquiring
fluorescence images is a particular concern because the fluorescence can fade
(Z).
The
availability of high-quality digital cameras has significantly changed how
photomicrographs are used in the microbiology laboratory. Today, it is not
unusual for digital photomicrographs to be shared with experts via the
Internet. This process significantly extends the capabilities of the on-site
microbiologist and can enhance patient care. Microscope-based digital cameras
and video systems are also used to perform "plate rounds in remote
hospitals and clinics within a multihospital system.
Newer Internet
technologies involving robotic microscopes and high-resolution video systems
now allow microbiologists to change the focus and change slide positioning of a
microscope located anywhere in the world and view the resulting images on a
monitor in their office. The availability of digital photomicroscopy has
significantly enhanced the microbial identification process, and it has helped
to standardize microbe identification. A wide variety of microscopes are
currently available that have integrated camera systems and sophisticated light
metering and exposure controls. Accessory cameras are also available from a
large number of aftermarket manufacturers.
The performance and optical characteristics of
these systems are too diverse to discuss in a single chapter, and the camera
specifications necessary for digital microscopy will depend upon the type of
images that will be captured. Light sensitivity is a key element in any digital
camera, and camera
Manufacturers
utilize the film-equivalent International Organization for Standardization
(ISO) numbers to rate the light sensitivity of image sensors in digital
cameras. ISO numbers range from 80 to 3200 and the higher the ISO value, the
more sensitive the image sensor. Some digital cameras have manual or automatic
sensitive adjustments that can alter the image sensor so that the camera can be
used in low-light conditions. Unfortunately, this increased sensitivity is
usually achieved by amplifying the signal from the image sensor. This type of
Signal
amplification also increases the background levels and decreases image quality.
The "heart" of the digital camera is the image sensor, a silicon chip
that measures and captures light. Presently, there are two types of image
sensors, the complementary metal- oxide semiconductor (CMOS) and the charge-
coupled device (CCD) sensor. Consumer cameras typically use CMOS chips because
they are easier to manufacture. Many CMOS cameras can be used to capture images
generated during bright-field microscopy. CMOS cameras tend to be smaller and
less expensive than CCD cameras. In addition, CMOS cameras use less power, and
they have faster frame rates, fewer artifacts (smear and blooming) caused by
charge transfer between adjacent pixels, and the ability to include
"higher-level" camera functions such as image stabilization and
wireless control . The major disadvantage of the CMOS camera is its lower light
sensitivity.
CCD and
slow-scan CCD cameras, which were first developed for astronomy, are the
current cameras of choice for low-light-level fluorescence microscopy. However,
these cameras are much more expensive than CMOS cameras. Photographs are a
stern judge of microscopic quality, and the purchase of an expensive camera
system does not automatically confer the ability to produce high-quality
images. Optics, optical train alignment, and proper illumination are the most
important factors in the acquisition of high-quality photomicrographs. Optical
image deficiencies as evidenced by chromatic variance and poor image clarity
are more noticeable when using a digital camera than when using 35-mm film.
Color photography can be especially demanding because specimens may appear
yellow or blue under tungsten halogen (3,200-K color temperature) light depending
upon whether the lamp voltage is above or below the recommended g-volt setting.
Photographs will also appear blue if the daylight blue filter is not removed
from the light path. Photographs can also appear yellow when tungsten halogen
(3,200-K) illumination is used in conjunction with daylight (5,500-K) film or
digital cameras designed for daylight photography. Under these conditions, a
Kodak 80A (3,200-K to 5,500- K) color conversion filter should be placed in the
light path to achieve the proper 5,500-K color temperature.
Not all
microbiologists can afford a microscope with an integrated camera system.
Simple eyepiece cameras can also be used to capture bright-field images. The
simplest configuration for eyepiece photography involves the use of a point-and-shoot
digital camera. Some improvisation may be necessary with this method because
few adapters are available for coupling a fixed-lens camera to the microscope
eyepiece. Instead, the microscopist can use a camera tripod or some other
support bracket to hold the camera in its proper position. Entry of stray light
can be minimized by using a piece of black polyvinyl chloride (PVC) pipe with
an appropriate diameter and a black camera cloth. During photography, the
camera lens system should be set to infinity focus (the default in fixed lens
cameras), and the lens should be positioned so that it is at the eye point
(focal point) of the eyepiece. The location of the eyepoint can be determined
by holding a piece of white paper just above the objective with the microscope
turned on and focused A bright circle of light will be projected onto the
paper. The eyepoint is the position where the light circle is smallest.
Photographs produced under these conditions are often acceptable, but they may
be dark. Cameras with adjustable aperture settings should be set to the largest
apefture value (smallest f-stop number) to maximize the amount of light
entering the camera. This method will also produce some chromatic aberration
(due to different lens correction factors) and vignetting (pipe view effect).
Another method for photomicroscopy is to use
the camera port on microscopes fitted with a trinocular head. Olympus and Nikon
have introduced adapters that allow their digital cameras to attach to the
camera tube of their microscopes. In addition, camera tube and eyepiece
adaptors for a number of digital cameras are available from Microscope Depot
(Tracy, CA). Photography under these conditions is best done using a camera
with through- the-lens exposure metering. These devices work well if the
exposure is no longer than several seconds or shorter than one-third of a
second Many of these cameras have built- in flashes that should be turned off
during photomicroscopy. These cameras may have problems with fluorescence microscopy
due to the extreme contrast of fluorescent images and the tendency of metering
systems to average exposure values over the entire field.
CARE
AND USE OF THE MICROSCOPE
Proper care
and maintenance of the microscope will prolong the usable life of the
instrument and allow for more accurate interpretation of microbiological
images. The microscope should be kept in a low-vibration, low-dust environment
to facilitate viewing and decrease damage to the optical systems. The optical
elements should be kept completely free of dust, dirt, oil, solvents, and any
other contaminants. Ideally, the microscope should be covered and the lamp
should be turned off when the microscope is not in use.
Do not touch
the optical surfaces with your fingers. Keep the lenses clean and be sure to
remove oil or mounting fluid from the objectives, condenser, and mechanical
stage after each session. Avoid dragging the high-dry objective through oil or
fluorescence mounting fluid. One way to avoid accidental contact with these
fluids is to place the high-dry objective and the oil immersion objective in
the nosepiece on opposite sides of the owl-power objective Lenses should be
dusted with residue-free compressed air and cleaned with lens paper and a
commercial lens cleaner that is approved by the microscope manufacturer.
Organic solvents such as alcohols and acetone should not be used on the lenses
because the solvent may dissolve the optical mounting cement. Unused spaces in
the nosepiece should be plugged and the eyepieces should remain installed at
all times to prevent introduction of dust into the body tube. The stage should
be cleaned regularly, and any spilled immersion oil must be removed or slides
will stick when they are moved across the stage. Spilled immersion oil also
collects dust and grit that can damage the optical and mechanical parts.
Microscopists should not attempt to remcwe or dissemble the objectives, as this
increases the potential for damage.
This is a job
that is best left to professionals. The gears and rackwork should be cleaned
and treated with new grease at intervals specified by the manufacturer. Do not
use light oil on the gears or bearing surfaces because this may cause the
condenser and stage to sink from their own weight. Periodic cleaning and
adjustment by a professional microscope repair person also help to extend the
usable life of the microscope.
ERGONOMICS
Peering into a
microscope eyepiece for long periods is not an activity for which the body is
well adapted. Microscope work requires the head and arms to be locked in a
forward position and inclined toward the microscope with rounded shoulders.
This unusual positioning is further exaggerated when the feet are placed on the
ring-style footrests that are common to many laboratory stools.
Poor posture and awkward positioning during
microscopy can cause pain or injury to the neck, wrists, back, shoulders, and
arms. In one regional survey of cytotechnologists, Kalavar and Hunting found
that 70.5% of respondents reported neck, shoulder, or upper back pain during
microscopy and 56% had an increased prevalence of hand or wrist symptoms.
Eyestrain and leg and foot discomfort have also been documented with long-term
microscope use. When using older microscopes, users often have their heads
inclined up to 450 from vertical and their upper backs may be inclined by as
much as 300. Even 300 inclinations of the head can produce significant muscle
contractions, fatigue, and pain For this reason, microscopists should be taught
to sit upright and hold their heads in neutral positions (3Z)• During
microscopy, the laboratorian should sit erect and maintain the natural curve of
the spine. The lower back and shoulder blades should be supported by the chair,
and a lumbar support cushion should be used if necessary. The legs and feet
should rest firmly on the floor or a footrest. The chair should have a pneumatic
height adjustment, and the seat should have a sloping front edge to prevent
undue pressure on the thighs. The backrest should be adjustable for both height
and angle, and the chair should have a five-pointed star base with caster
wheels. Knee spaces, which are often used for laboratory storage, should be
free from obstructions, and there should be a minimum of 2 inches of clearance
between the thigh and the bottom of the desk or counter. Obstructions that
prevent the microscopist from holding his or her shoulders perpendicular to the
ocular axis of themicroscope should be removed. The upper arms should be
perpendicular to the floor, with the elbows close to the body. The forearms
should be parallel with the floor, and the wrists should be straight.
The head
should be upright, and the neck should bend as little as possible, preferably
no more than 10 to 150. The eyepieces should be just below the eyes, and the
eyes should 00k downward at a 30 to 450 angle. The use of tilting microscope
heads can significantly improve the comfort of the microscopist. Repetitive
motions of the hands and the contact stress of arms resting on (the edge of) a
hard surface can cause pain and nerve injury, leading to repetitive stress
injuries and/or carpal tunnel syndrome. The use of padded arm rests can
moderate some of these problems. In addition, to prevent stiffening of the
muscles during microscopy, microscopes should not be placed under an air vent.
Most
laboratory microscopes are used by multiple individuals, and it is often a
challenge to find conditions or microscope configurations that satisfy
everyone. Some laboratories place microscopes on books or heavy,' blocks of
wood to accommodate taller microscopists. This configuration creates a number
of problems. If the microscope is raised to a sufficient height to prevent neck
flexion, users may be forced to bend their wrists into an unnatural position.
If the microscope is lowered to allow the forearms to remain parallel to the
floor, the neck is forced to bend. Lowering the chair to its lowest position
causes leg discomfort.
Individuals of
short stature may have to raise the chair to a level where their feet no longer
touch the floor. Foot rests can ameliorate this problem, but some individuals
may have insufficient space under the bench top to accommodate their legs. In
practice, most laboratories will elect to use a suboptimum, but workable,
microscope configuration that all users can employ. Under these conditions,
microscopists can reduce stress and fatigue by taking 1-minute
"microbreaks" every 10 to 15 minutes where they can stand, stretch,
and allow the eyes to focus at a distance. More expensive solutions, including
the use of ergonomic microscope tables that can be raised and lowered
electrically have been employed in some laboratories. Eye fatigue can be a
major problem for microscope users, especially if they have poor vision. The
diopter adjustment provided on most microscope eyepieces can be adjusted to
compensate for minor near- and farsightedness, thereby allowing users to remove
their glasses during microscope use. The diopter adjustments do not adjust for
astigmatism, and users with moderate to severe astigmatism should wear glasses
when using the microscope. Most microscope manufacturers now produce
high-eyepoint eyepieces that move the visual observation point further from the
eyepiece, thereby facilitating the use of glasses during microscopy. Ensuring
that the microscope images are as bright, sharp, and crisp as possible will
also help to reduce eye fatigue and associated headaches. The importance of
proper arrangement of the microscope and optical components cannot be
overstressed. Proper optical arrangement and the use of newer objectives with
higher NA values will produce brighter images and better resolution, which
eases the strain of searching for tiny specimen details.
The use of a
neutral blue (day light) filter during bright-field microscopy can also help to
lessen eyestrain when examining microbiological specimens. In the future, many
new microscopes will display the specimen image on a computer monitor. This
innovation could alleviate many of the eyestrain problems that develop during
extended microscope use.
Microscopes
are as different as the people who use them, and the previous comments should
not be construed as a prescription for alleviating strain or repetitive motion
injuries in every situation. When purchasing a microscope, every effort should
be made to allow microscopists to evaluate the new microscope under their
normal working conditions. Some microscopes will be comfortable for some users
and uncomfortable for others. In the long run, the feel and fit of the
microscope are just as important as the optical characteristics.
CONCLUSION
Advances in
the design, resolution, and ergonomics of modern microscopes have greatly
enhanced our ability to study and identify microorganisms. Microscopy still has
a central role in the detection of infectious agents despite high y publicized
advances in DNA and RNA detection systems. Microscopic examination of clinical
specimens provides a rapid and inexpensive "first pass" in the
detection and identification of infectious agents. Thus, clinical microscopy
will continue to be a core competency in clinical microbiology laboratories for
the foreseeable future.
No comments:
Post a Comment